Skip to main content
Search

* All fields must be filled out.

Send message!
Search
EHY-2030-Latest device of Oncotherm

EHY-2030

Overview

Our new, revolutionary loco-regional electro-hyperthermia device

EHY-2030 is our latest development in the treatment of loco-regional tumors. The newly designed device includes the Smart Electrode System (SES), the integrated Patient Management System (PMS-100) and a user-friendly touch screen display with full system control. The new RF generator with increased power has been developed with a new intelligently controlled step motor tuning system for rapid impedance matching to achieve faster tuning times. The device is only for clinical use.
 

Demo video

Characteristics

New shape and design
User friendly touch screen display with full system control
New automatic controlled step motor tuning system for rapid impedance matching to achieve faster tuning times
Newly developed RF generator with increased power
Robotically controlled electrode arm with physician feedback to easily and accurately position the smart electrode
Changeable stretching textile electrode for the smart electrode system and bed
Hand-held emergency stop for patients

Indications

The EHY-2030 device is intended for the adjuvant treatment of glioblastoma in combination regimen with chemotherapy and/or radiation therapy.

Contraindications

Cannot be used when the patient is under deep-sedation or anesthesia (missing thermal sensitivity). Application of analgesics in the treated area is prohibited.
Do not use on patients with reduced ability to pain perception.
Cannot be used when the patient is unconscious.
Cannot be used on patient with reduced ability to communicate.
Do not use on patients who have implanted active medical device in any anatomical location (e.g. cardiac pacemaker, implanted erectile function stimulator, etc.)
Do not use on patients who have implanted any medical device in the neck or head (e.g. implanted deep brain stimulator, implanted hearing-aids, etc.)
Do not use on patients with non-removable object on the neck or head.
Do not use on patients with open wound, hemorrhagic syndrome or unformed postoperative suture on head or neck.
Do not apply on pregnant patients.
Do not apply on patients with Karnofsky performance index < 30%.

Technical specifications

Main voltage
100 - 230 VAC 50/60 Hz
Power input
max. 800 W
RF power output
max. 350 W(eff) / 800 W(pp)
Nominal load
50 Ohm
Output carrier frequency
13,56 Mhz
Output modulating frequency
Pink noise
Weight
200 kg
Dimensions (HxLxW)
1320 x 2500 x 1150 mm
Temperature
+15°C - +30°C
Relative air humidity
20% - 60% (non-condensing)
Air pressure
800 - 1060 hPa